Tandem dye naming cheat sheet: APC, PE, and PerCP variants across vendors

APC-Fire 750, APC-Cy7, APC-H7, and APC-eFluor 780 are all red-laser tandems emitting near 775–785 nm, but they come from different vendors and have different stability profiles. This page maps every major PE, APC, and PerCP tandem variant to its emission peak and cross-vendor equivalents, so you can find substitutes or decode unfamiliar names quickly.

What this page covers

Commercial tandem naming by donor (PE, APC, PerCP) with emission peaks and cross-vendor equivalents.

Why the same dye has so many names

Tandem fluorophores are proprietary; each vendor independently conjugates the same donor dye to different acceptors, or to the same acceptor using a different linker chemistry. The result is a proliferation of trade names that describe essentially the same spectral position but carry different part numbers, pricing, and stability characteristics.

For example: PE-Cy7 (BD, BioLegend, Invitrogen) and PE/Fire 810 (BioLegend) both use PE as the donor and a Cy7-like acceptor, but Fire 810 emits ~25 nm longer, which means they are not drop-in substitutes on a conventional instrument, even though the names sound like the same class of dye.

Conversely, APC-Cy7 (BioLegend) and APC-H7 (BD) are close enough to be interchangeable on most instruments, but APC-H7 uses a proprietary stabilized linker that gives more consistent lot-to-lot performance.

The emission peak is the ground truth. When in doubt, compare spectra rather than names.

Tandem variant reference

Select a donor dye to see all commercial tandem variants sorted by emission peak. Cross-vendor equivalents are grouped together so you can find a substitute at a glance.

Blue 488 nm or YG 561 nm excitation
NameEm. peakVendor(s)Cross-vendor equivalentsNotes

PE-eFluor 610

610 nmInvitrogen (eBioscience)
  • PE-Dazzle 594 (BioLegend)
  • PE-CF594 (BD)
eBioscience branding. Spectrally very close to PE-Dazzle 594. Check emission spectrum before assuming direct substitutability.

PE-Dazzle 594

612 nmBioLegend
  • PE-CF594 (BD)
  • PE-Texas Red (BD, Invitrogen)
BioLegend's version of PE-CF594. Very similar emission; often interchangeable on the same filter set.

PE-CF594

612 nmBD Biosciences
  • PE-Dazzle 594 (BioLegend)
BD's Texas Red-based tandem. Excellent brightness and stability compared to older PE-Texas Red. Good first choice in the 610 nm window.

PE-Texas Red

615 nmBD / Invitrogen
  • PE-CF594 (BD)
  • PE-Dazzle 594 (BioLegend)
Older tandem; less stable than CF594. Most labs have migrated to PE-CF594 or PE-Dazzle 594 for better lot-to-lot consistency.

PE/Fire 640

640 nmBioLegendBioLegend Fire series tandem at 640 nm. Fills the gap between PE-CF594 and PE-Cy5. Less common; verify filter availability on your instrument.

PE-Cy5

667 nmBD / BioLegend / Invitrogen
  • PE-Vio 670 (Miltenyi)
Older tandem with known instability. Susceptible to spectral spillover into PE channel upon degradation. PE-Vio 670 offers improved stability as a Miltenyi alternative.

PE-Vio 670

670 nmMiltenyi Biotec
  • PE-Cy5 (BD, BioLegend)
Miltenyi's RE AffinityDye-based alternative to PE-Cy5. Generally more photostable than Cy5-based conjugates.

PE-Cy5.5

695 nmBD / BioLegend / Invitrogen
  • PE/Fire 700 (BioLegend), slightly longer
Prone to FRET instability like PE-Cy5. Lot-match your single-stain controls. Consider PE/Fire 700 for better stability.

PE/Fire 700

700 nmBioLegend
  • PE-Cy5.5, similar channel
BioLegend Fire tandem with better stability than PE-Cy5.5. A practical upgrade if you are experiencing lot-to-lot variability with PE-Cy5.5.

PE/Fire 744

744 nmBioLegendFills spectral gap between PE-Cy5.5 and PE-Cy7. Useful on full-spectrum instruments. Cross-vendor equivalent is uncommon.

PE-Cy7

785 nmBD / BioLegend / Invitrogen
  • PE/Fire 810 (BioLegend), slightly longer
Most widely used PE tandem but highest instability risk. Light and oxygen exposure degrade signal quickly. Always run fresh single-stain controls.

PE/Fire 810

810 nmBioLegend
  • PE-Cy7, shorter emission
Longer-emission Fire tandem; useful on spectral instruments to push PE-based detection further into the near-IR.

Similar-sounding dyes: know the differences

Fluorochrome names often borrow color words that overlap across brands, creating confusion about whether two dyes are interchangeable. These groups are particularly prone to mix-ups.

Blue emission~450 nm · Violet/UV laser
DyeMakerEm. peakKey note
Pacific BlueInvitrogen~450 nmUV or violet excitation; original 'Pacific' dye
BV421BD Biosciences~421 nmViolet 405 nm; brighter than Pacific Blue
VioBlueMiltenyi~450 nmViolet 405 nm
V450BD Biosciences~448 nmOlder BD violet dye; largely replaced by BV421
Atlantic BlueInvitrogen~440 nmUV 355 nm excitation; not violet!

Watch out

Green emission~510 nm · Violet laser
DyeMakerEm. peakKey note
BV510BD Biosciences~510 nmViolet 405 nm; excellent for viability or dump channels
VioGreenMiltenyi~530 nmViolet 405 nm; slightly longer emission than BV510
V500BD Biosciences~500 nmViolet 405 nm; older version of BV510
Pacific OrangeInvitrogen~551 nmViolet 405 nm; emission at ~551 nm is further out than BV510/VioGreen; confirm your detector before substituting
Krome OrangeBeckman Coulter~528 nmViolet 405 nm; unique to Beckman Coulter instruments

Watch out

Pacific Orange emits at ~551 nm, further than BV510 (~510 nm) or VioGreen (~530 nm). All three use the violet 405 nm laser, but they do not share the same detector. Confirm your instrument's filter configuration before treating them as interchangeable.

Before you substitute one vendor for another

Cross-vendor substitution is usually possible within the same emission group, but it requires verification. Run through this checklist before swapping.

  • Confirm the substitute shares the same excitation laser and filter configuration on your instrument.
  • Compare emission spectra; even a 15 nm shift can require a compensation update.
  • Use a spectrum viewer (BD, BioLegend, FPbase) to simulate the new dye in your full panel before ordering.
  • Order a test lot before committing to a large purchase; run a side-by-side titration with the original.
  • Update your compensation matrix using single-stain controls made with the new substitute, not the old one.
  • If the experiment is longitudinal, document the substitution in your methods; it's a reagent change.

Same peak, different spectrum

Two tandems with the same emission peak can have different spectrums. This matters much more in spectral cytometry than conventional cytometry, but check every detail and assume nothing.

References

  • [1]Roederer M. Spectral compensation for flow cytometry: visualization artifacts, limitations, and caveats. Cytometry. 2001;45(3):194–205.
  • [2]Tung JW, et al. Modern flow cytometry: a practical approach. Clin Lab Med. 2007;27(3):453–68.
  • [3]Baumgarth N, Roederer M. A practical approach to multicolor flow cytometry for immunophenotyping. J Immunol Methods. 2000;243(1-2):77–97.