Skip to content
Back to TiterPubs

CD45RO · BV480

Dataset ID: YQKC5784 • DOI: 10.64157/YQKC5784

Analysis software: ACTA 3.6.11

Contributed together with

Same submission, acquired 2026-10-20: 6 datasets, one per reagent. The full bundle and the report are the run's, the same on all 6 pages; the FCS files are this reagent's. Cite each dataset you use.

Figures

Stain index. Stain index of live CD14/CD16-negative (DUMP-) cells at each stain volume, with ACTA's Michaelis-Menten fit. One panel per reagent in the run (top row CD127, CD19, CD27, CD45RA; bottom row CD45RO, CD62L): CD45RO is the first panel of the bottom row. For CD45RO, the fit (R² 0.9997) puts 90% of the fitted maximum at 59.7 µL, beyond the highest volume tested (10 µL); the highest stain index, 108, is at 10 µL, the highest volume tested.
Saturation. Share of live CD14/CD16-negative (DUMP-) cells in the positive gate at each stain volume, with the maximum ACTA marks. One panel per reagent in the run (top row CD127, CD19, CD27, CD45RA; bottom row CD45RO, CD62L): CD45RO is the first panel of the bottom row. For CD45RO, the highest share, 75.8% of live DUMP- cells, is at 10 µL.
Concatenated events. CD45RO fluorescence (BV480-A) of live CD14/CD16-negative (DUMP-) cells at each stain volume, from the 0 µL co-stain control to 10 µL, each well's events side by side (up to 20,000 per file). One panel per reagent in the run (top row CD127, CD19, CD27, CD45RA; bottom row CD45RO, CD62L): CD45RO is the first panel of the bottom row.
Gating scheme. The run's gating hierarchy, one tree for all six reagents, each gate coloured by the method that drew it: cells, single cells and live cells, then a combined CD14/CD16 dump gate (BUV615) on live cells, and on the dump-negative (DUMP-) cells the reagent's positive and negative gates.

Experiment Information

Name
Jatin Sharma
Month of analysis
November
Year of analysis
2026
Collection Software
SpectroFlo
Analysis Software
ACTA
Plug-ins, other analysis tools
openCyto 2.22.0; flowCore 2.22.1; AutoSpectral 1.6.2 (unmixing)
References
https://doi.org/10.5281/zenodo.24181307
Antigen
CD45RO
Clone
UCHL1
Conjugate
BV480
Target Species
Human
Antibody host species
Mouse
Clonality
monoclonal
Supplier
BD Biosciences
Catalog #
566143
Lot #
5331625
Other antibodies in tube
CD14 BUV615; CD16 BUV615; CD3 BUV805; CD4 BUV496; CD8 Alexa Fluor 532; Live Dead eFluor 780 (LD) (Viability) efluor780 · see the table
Instrument Manufacturer
Cytek
Instrumentation
Aurora
Lasers (nm)
355, 405, 488, 561, 640
Incubation Time
30 minutes
Incubation temperature (°C)
4
Incubation volume (µL)
100
Gating
live single cells ·
Tissue
blood
Cell type
mononuclear cells
Cell count
1000000
Fresh or Frozen
Frozen, thawed and incubated with antibody
Live or Fixed
Live at the time of antibody incubation
Fixed after antibody incubation?
Yes, antibody then fixed
Acquisition date
2026-10-20

Contact the contributor via MetrdyWe pass your message on; contributors' addresses are not published.

Other reagents in the tube

The panel the titrated reagent was stained with. Base reagents are fixed in the panel; Variable is the reagent this dataset titrates.

AntigenFluorochromeCloneVendorCatalog #Role
CD14BUV61561D3Invitrogen366-0149-42Base
CD16BUV6153G8BD OptiBuild751572Base
CD3BUV805SK7BD612893Base
CD4BUV496SK3BD612936Base
CD8Alexa Fluor 532RPA-T8Thermo58-0088-42Base
Live Dead eFluor 780 (LD) (Viability)efluor780Not givenTS65-0865-14Base
CD45ROBV480UCHL1BD Biosciences566143Variable

A titer holds for the conditions it was run under: cells, instrument and staining conditions all change it. Metrdy checks each submission for completeness before publishing it, but does not validate the results.

ACTA downloads

Full analysis bundle: re-runnable

FCS files

The 11 files

CD45RO_E1.fcs, CD45RO_E2.fcs, CD45RO_E3.fcs, CD45RO_E4.fcs, CD45RO_E5.fcs, CD45RO_E6.fcs, CD45RO_E7.fcs, CD45RO_E8.fcs, CD45RO_E9.fcs, CD45RO_E10.fcs, CD45RO_E11.fcs

FCS files (.zip, 11 files, 198 MB)

Full ACTA bundle

Everything ACTA wrote for the run, the same file on all six reagents' pages: a filled TiterPubs template per reagent, the export and instructions workbooks, the panel inventory, the version files, the report, every plot and all 66 FCS files.

Full ACTA bundle (.zip, 1.3 GB)

ACTA report

ACTA's report on the whole run, all six reagents: its MiFlowCyt metadata, the plate map, the QC checks, every plot, the run's settings and the software versions.

Open in a new tabDownload (35.6 MB)

DOI

https://doi.org/10.64157/YQKC5784

Licence

CC BY 4.0Anyone may share and adapt this dataset, crediting the contributor and citing it.

Citation

Sharma, J. (2026). CD45RO BV480 Titration Dataset (TiterPubs YQKC5784) [Data set]. Metrdy. https://doi.org/10.64157/YQKC5784

Analysis software

ACTA 3.6.11 (Anybody Can Titrate Antibody)

Open-source (AGPL-3.0-only), by Jatin Sharma.Copyright (C) 2026 Lyell Immunopharma, Inc.

Bundle and templates written by ACTA's TiterPubs export.

How to cite ACTA

Sharma J. ACTA — Anybody Can Titrate Antibody [software]. Version 3.6.11. Lyell Immunopharma, Inc.; 2026. https://github.com/jatin199517/acta-public. https://doi.org/10.5281/zenodo.24181307

Until the ACTA paper is published, its author asks for ACTA to be cited by its GitHub repository and the version used, and by that version's DOI.

Named because the contributor analysed the data with it. Metrdy did not write ACTA and does not validate its results.

Reproduce this analysis

The bundle holds the instructions workbook that drove the run and the FCS files ACTA read. Install the ACTA version named above, unzip the bundle into an empty folder and run run_acta() on that folder.

remotes::install_github("jatin199517/acta-public@v3.6.11")library(ACTA)r <- run_acta("path/to/the/unzipped/bundle")

ACTA read every event of each file (Downsample_to is blank in the Info sheet; acta_version.tsv records seed 123): 24,354,336 events across the run's 66 files. Its report says the identity matrix was applied and the data were not compensated; the Info sheet's compensation setting is blank and the bundle holds no compensation file. The FCS files are AutoSpectral 1.6.2 (OLS) unmixing output, as their keywords record.

Used this titer? Tell other labs how it went. Download the report form, then upload it on the Contribute tab or email it to us. We review every report and add it here, signed with your name.

How community reports workReport a problem privately