ACTA (Anybody Can Titrate Antibody) is a free, open-source tool that turns FCS files into a stain index and a fitted saturation curve. Try its core calculation below on ACTA’s own validation data by adjusting the blue sliders, then see how ACTA results become published datasets on TiterPubs.

Describes ACTA 3.6.11 (released 12 November 2026)Last checked 19 November 20268 min read, plus the demo

ACTA logo

About this page

ACTA was written by Jatin Sharma at Lyell Immunopharma, Inc., which holds its copyright, and is published under the GNU AGPL, version 3 only (AGPL-3.0-only) [1]. Metrdy did not build ACTA and is not affiliated with Lyell, and nobody was paid for this page. Metrdy runs TiterPubs.

ACTA logo © 2026 Lyell Immunopharma, Inc., used with permission.

ACTA streamlines titration analysis

ACTA automates the otherwise tedious, subjective gating and calculations of antibody titrations. You give ACTA a folder of FCS files and one Excel workbook; the gating hierarchy and the run settings live in that workbook, not in code. It handles combinatorial titrations too, where several reagents share the same tube [2].

ACTA is an R package with a Shiny app that runs the whole workflow; using it takes no R experience. ACTA calls the workbook its instructions workbook, and the workbook’s Info sheet holds the run settings. For each titration it gates every well by the hierarchy in the workbook (for example cells, single cells and live cells, then the titrated reagent), using gating methods run through openCyto [3]. It then computes the stain index at each dose [4] and fits a saturation curve, SI = Vmax·dose / (Km + dose) [5]. From these it suggests two titers: the dose with the highest stain index, and the dose where the fitted curve reaches 90% of its plateau. Each run ends with plots, an Excel export and, optionally, a PDF report. A separate step builds a dashboard across your runs.

Getting started with ACTA:

Live demo

Stain index and titer on ACTA’s validation data

This demo re-creates ACTA’s stain-index and plateau-fit steps in your browser, on OQ_Test1, one of the validation datasets that ship with ACTA: a streptavidin-PE titration acquired on a Cytek Aurora. The same data, analysed by ACTA 3.6.11, is published on TiterPubs as doi:10.64157/DVFD8329. The events were gated to live single cells beforehand, with gates that approximate ACTA’s (the list under the demo says how). The demo shows one of ACTA’s six QC checks, the zero-dose check; ACTA itself runs all six and writes the report and the provenance files.

The live demo loads as you scroll to it.

Reading the result

Where a gate sits changes the stain index at that dose, and with it the titer. ACTA sets each gate by the same rule every time. Two people who analyse the same files with the same workbook get the same answer. For this run, ACTA’s rule splits each dose in two but never sets the gate below the zero-dose well’s 99th percentile. “One gate for all doses” puts every gate at that 99th percentile instead.

ACTA suggests two titers. One is the dose with the highest stain index [4]. The other is the dose where the fitted curve reaches 90% of its plateau, which is 9 × Km [5]. The two can be many-fold apart. ACTA leaves the choice to the analyst, and Bonilla et al. advise at least twice the 90% dose [6].

Look at R² before using the 90% titer. In this series the stain index jumps about five-fold between 0.02 and 0.04 µL, then climbs slowly to the top dose. A saturating curve follows that shape only roughly: ACTA’s own fit has R² 0.879. ACTA’s app offers the 90% titer as a one-click choice only when R² is at least 0.9 and the dose lies inside the doses tested, so it does not offer it for this series; the export includes the number either way. A series whose stain index falls again at the top doses fits poorly too.

How ACTA checks its results

Three diagnostic cases ship with ACTA as its validation data. They test the software, and they are worked examples of a correct instructions workbook: OQ_Test1, one reagent on a Cytek Aurora (the demo’s data); OQ_Test2, three titrations on a Bio-Rad ZE5 with a compensation matrix; and OQ_Test3, a combinatorial case on the Aurora, with two reagents stained together in one set of FCS files. To validate an installation, ACTA’s README suggests running actaOQRun() on them; the Diagnostics section of ACTA’s app runs them too. Each case runs the whole analysis and returns PASS, PASS (with notes) or FAIL. A pass means the run finished and returned the values expected for that case: it checks the software, not the reagents.

Every titration also goes through six QC checks, and each verdict is written to the export. The median forward scatter of the Cells gate must be more than half that of all events. In the worst well there must be more than 50% live cells and more than 80% single cells. The zero-dose control must have the lowest stain index in the series (a tie passes), and be less than 5% positive. And every gated population in every well needs at least 100 events, except the positive population of the zero-dose control.

Two of ACTA’s gating methods, flowClust and flowMeans (run through openCyto [3]), start from random values. ACTA sets the random seed to 123 at the start of each antibody’s analysis. ACTA’s internals document records a test of this: the pipeline was run in two separate R processes, and all 36 gate percentages matched to ten decimal places.

Each run also writes acta_version.tsv and package_versions.tsv next to its results. They record the seed, the script’s version, the pipeline folder that ran and the version of every package the run could have used.

From ACTA’s README, docs/INTERNALS.md and R/ACTA_Functions.R at version 3.6.11.

Publishing ACTA results on TiterPubs

TiterPubs is Metrdy’s free library of titration datasets, and browsing it needs no account. Each dataset has a Crossref DOI under Metrdy’s prefix, 10.64157, and names its contributor, with their ORCID iD when they give one. New datasets are shared under CC BY 4.0, which allows reuse with attribution.

Two examples of ACTA results on TiterPubs: the demo’s own data, published as streptavidin-PE (doi:10.64157/DVFD8329); and one run of three reagents, published as three linked datasets with a DOI each: CD19 (doi:10.64157/UQTO5148), CD56 (doi:10.64157/LYKL0014) and CD7 (doi:10.64157/PKER2786).

Publishing on TiterPubs does not mean Metrdy has validated a titer. The titer is the contributor’s own result.

ACTA’s TiterPubs export

Since version 3.6.0, ACTA can write a TiterPubs bundle: one zip file that holds everything TiterPubs needs to publish a run. ACTA runs on your own computer, and nothing reaches Metrdy unless you upload it. These steps follow ACTA 3.6.11. Use 3.6.11 or later: earlier releases also write a bundle, but their scrub does not blank the SOP field.

Making a bundle

  1. In the Info sheet of the run’s instructions workbook, set titerpubs_export to TRUE. It is FALSE by default, so a run writes no bundle unless you ask for one. Turn it on only for a run you mean to publish under CC BY 4.0.
  2. In the same sheet, fill in the TiterPubs, Protocol and Equipment rows: the name and ORCID iD to credit, and details such as the QC beads, the acquisition software, the target species, the incubation time and temperature, the instrument’s make and model, how the cells were prepared and the staining protocol. ACTA fills each reagent’s TiterPubs template from these rows and the plate layout. The blank workbook comes with example values filled in: the name Wile-E, Acme Corporation in Organization/Affiliation, an example address in Submitter_Email, and example links and settings in other rows. Check every row, because what you leave in is published. Clear Submitter_Email and leave it empty (step 5 says why).
  3. In the plate layout (the Layout_Plate sheet), give every reagent its catalogue number and vendor (Cat_Num and Vendor). They say which reagent a dataset describes, and ACTA writes no bundle while one is missing.
  4. Decide whether to set scrub_confidential to TRUE (below). Then run ACTA as usual, with the PDF report on: ACTA builds the bundle only when the run makes its report. The report needs pandoc and a TeX engine, and the bundle needs the pdftools R package. ACTA’s Setup.R installs the TeX engine and pdftools but not pandoc; ACTA’s installation notes on GitHub say how to get it. The bundle is a zip file written with the run’s other results, and its name ends in _ACTA_TiterPubs.zip. If the report fails, ACTA still finishes the analysis but writes no bundle, and its log says why.
  5. Upload the bundle, still zipped, on the Contribute tab of TiterPubs, and give your email address in the upload form (or sign in to Metrdy first). ACTA removes Submitter_Email from every bundle, so the form is how we reach you. One upload can be up to 3 GB (50 files), in one file or several; for a larger bundle, email us a share link instead (the Contribute tab gives the address).
  6. Metrdy staff check the submission for completeness before publishing it.

A run can titrate one reagent or several, and ACTA makes one bundle per run. Each reagent becomes its own TiterPubs dataset, with its own DOI, linked to the other reagents from the same run.

Definition of a bundle

  • one filled TiterPubs template per reagent;
  • the export workbook, which links each FCS file to its numbers;
  • the instructions workbook and the FCS files the run analysed;
  • the panel inventory, which lists the other reagents in the tube (the co-stain panel), and the compensation matrix, when the run used them;
  • the run’s report as a PDF, and the plots;
  • the run’s version files, acta_version.tsv and package_versions.tsv.

With these files, anyone can run the same analysis again: install the same ACTA version, unzip the bundle into an empty folder and run ACTA on it. The dataset pages give the steps.

Exports protect your data

A bundle is published with its datasets for anyone to download, so check what it carries before you send it. On every bundle, ACTA removes the instrument’s serial number (from the run’s own export and report too) and blanks Submitter_Email in the bundle’s copies of the instructions and export workbooks.

scrub_confidential is optional and FALSE by default. Set to TRUE, it puts placeholders in place of:

  • the operator’s name (JD);
  • the electronic lab notebook (ELN) ID (EXP26000001, unless you choose your own, below);
  • the ELN name (ELN);
  • the co-stain panel’s internal name (Iteration1, Iteration2 and so on);
  • the operator, project and institution keywords in the bundle’s copies of the FCS files.

It also blanks the SOP field. Left FALSE, ACTA publishes all of these as written, the SOP included. The scrub changes the run’s own report and export as well as the bundle: for an internal copy with the real names, run ACTA again with scrub_confidential set to FALSE. If a value the scrub replaces also appears in another column of the plate layout, ACTA writes no bundle and says why.

With the scrub on, every run publishes the same ELN ID. To tell related runs apart, add a row named scrub_eln_placeholder to the Info sheet, with the ID to publish in its Value column (letters, digits, dot, underscore and hyphen only). Do not use the real ELN ID: ACTA would publish it unchanged.

Free text is published as you wrote it: the cell type, sample details, references and the like. Keep donor and other confidential IDs out of it. File names are published too, and the scrub renames a bundled FCS file only when its name holds a value the scrub replaces, so keep them out of the names of the FCS files, the instructions workbook and the run folder too (the report prints the workbook’s name).

Results from other tools

No export in your analysis tool? Download the TiterPubs template from the Contribute tab, fill in one per reagent, and upload it with your FCS files in the same way. An ACTA older than 3.6.0 has no export: update ACTA and run the analysis again.

Share an ACTA titration

Upload it on the TiterPubs Contribute tab, or send a share link to the address given there. Each published dataset gets its own DOI.

References

  1. 1.

    Sharma J. ACTA — Anybody Can Titrate Antibody [software]. Version 3.6.11. Lyell Immunopharma, Inc.; 2026. Available from: https://github.com/jatin199517/acta-public . doi:10.5281/zenodo.24181307 . The tool this page describes. Until the ACTA paper is published, cite this repository with the version you used and that version’s DOI.

  2. 2.

    Burn OK, Mair F, Ferrer-Font L. Combinatorial antibody titrations for high-parameter flow cytometry. Cytometry A. 2024;105(5):388–393. doi:10.1002/cyto.a.24828 . The combinatorial approach ACTA supports.

  3. 3.

    Finak G, Frelinger J, Jiang W, et al. OpenCyto: an open source infrastructure for scalable, robust, reproducible, and automated, end-to-end flow cytometry data analysis. PLoS Comput Biol. 2014;10(8):e1003806. doi:10.1371/journal.pcbi.1003806 . The gating framework ACTA builds on.

  4. 4.

    Maecker HT, Frey T, Nomura LE, Trotter J. Selecting fluorochrome conjugates for maximum sensitivity. Cytometry A. 2004;62(2):169–173. doi:10.1002/cyto.a.20092 . The stain-index definition ACTA uses.

  5. 5.

    Johnson KA, Goody RS. The original Michaelis constant: translation of the 1913 Michaelis–Menten paper. Biochemistry. 2011;50(39):8264–8269. doi:10.1021/bi201284u . The curve form ACTA fits, used as an empirical saturation curve.

  6. 6.

    Bonilla DL, Paul A, Gil-Pulido J, Park LM, Jaimes MC. The power of reagent titration in flow cytometry. Cells. 2024;13(20):1677. doi:10.3390/cells13201677 . Applies the Michaelis–Menten form to antibody binding. Section 2.6 advises a titer at least 2-fold above the concentration that gives 90% of saturating staining.